Section 1
Specifications
All specifications are referred to plate voltage unless otherwise specified.
Charger ±1,100 volts (minimum), selectable polarity
Fieldmeter
Range: 0 to ±1,300 volts on display
0 to ±1,800 volts at analog output
1 volt resolution
Accuracy: ±2% of reading, ±2 volts, + zero offset, typical
Analog Output: 1/1000
th
of the plate voltage, ±3%, typical
Bandwidth: 6 Hz
Timer
Range: 0.1 second to 999.9 seconds
Accuracy: 0.1 second
Resolution: 0.1 second
Trip Points
Start Voltage: 1000 volts
Stop Voltage: Settable from 10 volts to 990 volts in 1-volt increments
Charge Plate
Size: 6 inches x 6 inches (15 cm x 15 cm)
Capacitance: 20 pF ±10%
Self-Discharge: Less than 100 volts within 5 minutes (<60% R.H.)
Power Requirements: 100 VAC to 220 VAC nominal line voltage; less than 5 watts
Battery Life: >8 hours per charge
Dimensions: 6-3/8” W x 6” H x 8-3/4” D (162mm x 152.4 mm x 223 mm)
Weight: 5.2 pounds (2.4 kg)